Light
Beams for Safety continued
It is our understanding that the OSHA regulation of using light
beams as personnel exclusion safety devices requires that
the light beam be sufficiently far from the machine to ensure
that the machine is fully stopped before anyone could be
hit by the rotating vessel. Specifically we have been told
that the separation has to allow for a distance of 66" per
second from when the beam is first interrupted to when the
blender
is fully stopped. The time must include the response time
of the light beam which can vary with manufacturers from
0.025 seconds upward.
As Gemco units use pneumatic brakes to provide a soft stop
and not impact the gears, our actual stopping times vary from
machine to machine based upon vessel RPM, capacity, shape,
unit fill level, certain options such as a Gemcomatic drum
loader and material density. We therefore can not provide the
actual stopping time for our units.
The time for large units can however be a full second or more.
This means that the light curtain would have to be positioned
more than 66" back from the extreme limit of rotation
to be compliant. Often such distances are not available within
the blending rooms so actual physical barriers are used instead.
The typical installation uses an interlock on the room door
to accomplish this goal. Some installations then go beyond
the door interlock and use a light curtain installed parallel
to the floor, perhaps an inch off the floor, to ensure that
no personnel stayed in the room once the door was closed.
Air System PMC continued
The most common inspections are daily PMCs, and these are simple
visual and mechanical checks that take no more than few minutes
to perform. A PMC vital to your Gemco equipment's continued good
heath is the Air System Check.
To ensure proper operation of your
GEMCO pressurized air is required. This air usually is piped to
the unit and connected to a filter, regulator and lubricator, mounted
on the equipment frame. The lubricator outlet is connected
to the remainder of the air system and the Main Air Pressure
Switch.
You
must
ensure
that there
is at least 85psi air supplied to the unit at all times through
a half-inch minimum pipe or hose to allow for the volume of air
required to operate the unit.
Inspect the lubricator and make sure
it is at least half full of lightweight 10 wt oil.
check
the o-ring seal for cuts or wear.
inspect the bowl itself look
for any small cracks or chips replace as required.
Inspect the
regulator looking for leaks or wear, broken parts. Inspect the
gauge face lens for clearness and replace parts as required.
Inspect
the filter
assembly,
remove the bowl and clean it as needed.
check to see if the drain
device is operating; look for cracking of the bowl or o-ring problems
on the drain valve, replace parts as required.
Check the rest of the air system for leaks, both while the unit
is stationary and rotating.
This simple check should take no more
than 15 minutes. It should be done at the start of every shift
if you run more than one shift, but it should be done at least
once a day.
Gemco Introduces New Lab Blender
The new Gemco cantilever solids type lab blender design includes
many new features. The motor and controls are enclosed in a
NEMA 4X stainless steel electrical cabinet. The safety gates
are built in and internally interlocked. Due to the wide range
of vessel sizes the unit is designed to accommodate, including
small vessels for 100 grams of powder or less, the
design includes a variable speed drive to enable the smallest
vessels to be run at the proper (higher) speeds required for
a proper blend as well as scale-up calculations when the quantities
increase.
|